ASTM D1401, ISO 6614

The adem by GreenLab for CANNON is an automated demulsibility tester that represents the next generation in ISO 6614 and ASTM D1401 method compliance. It features a unique design with PC control enabling fully automated operation and evaluation of measurement results. Incorporates state-of-the-art image processing and evaluating algorithm.

Common Applications

  • Lubricants and turbine oils
  • Compressor oils and engine oils
  • Synthetic fluids used in hydraulic and mechanic systems
  • Marine fuels/lubricants

adem Informational Video

State-of-the-Art Image Processing & Evaluating

  • 5 megapixel CCD camera
  • Measurement results presented in graphical or tabular format
  • Real time view of the sample compartment

Fully Automated and Adjustable Testing

  • Lower cost of operator error & analysis
  • Faster sample throughput, low cost of operation
  • Adjust, save and storage options of different method parameters
  • Storage capacity of over 10.000 measurement results
  • User-adjustable parameters for custom methods including custom stirring, temperature, measurement time and more
  • Bubble elimination function

Long Lasting and Safe Design

  • Long-life unique LED lightsource and other components
  • Low maintenance requirements
  • Multiple safety features, including stirring sensors, overheating, and bath level
  • Stable test-tube positioning
  • Stable bath temperature and bath discharge valve

Dimensions (W × D × H) 70 cm × 64 cm × 108 cm (14.3in × 26in × 28.5 in)
Weight 120 kg (140lb)
Bath Volume 18 L
Automated Sample Capacity 6 sample test tube holders
Stirrer Speed

500 to 1600 RPM (freely adjustable by user)

Stability of Stirring Speed ≤ ± 15 RPM @ 1500 RPM
Temperature Range & Accuracy Ambient to 90 °C ± 0.01 °C
Bath Temperature Stability < ± 0.01 °C
Power Supply 110 - 230 AC (VA)
Measurement Results Printed in graphs or tabular format on external printer
Control Software Windows®-based software compatible with Windows 7 or higher
Data Output USB or RS232 output

 