CCS-2100 Automated Cold-Cranking Simulator

For Apparent Viscosity of Engine Oils from –35 °C to –5 °C
ASTM D5293, SAE J300

The automated CCS-2100 is a cold-cranking simulator for measuring apparent viscosity of engine oils from –35 °C to –5 °C. Automated sample loading, operation and solvent-free cleaning permit unattended processing of up to 30 samples at one time.

Common Applications

  • Engine oils
  • Lubricating oils
  • Base stocks

Meets all ASTM D5293 and SAE J300 requirements and precision specifications

  • Instrument temperature range: –35 °C to –5 °C (± 0.05 °C)
  • Viscosity range: 900 mPa⋅s (cP) to 25,000 mPa⋅s (cP)

Fully automated operation

  • Automated sample loading and computer controlled testing of up to 30 samples at one time
  • Unattended operation reduces operator-to-operator variability for enhanced repeatability and reproducibility
  • Proprietary software automatically calculates and records sample viscosities based on test data and stored rotor/stator calibration information
  • The system automatically purges the previously measured sample with a portion of the next sample prior to viscosity measurment

Outstanding measurement control

  • Features a patented, thermoelectrically-cooled rotor/stator for outstanding temperature management
  • Rotor speed is automatically measured by a high resolution digital encoder

Reliable, convenient performance

  • Proven CANNON® reliability and outstanding support
  • A thermoelectric sample warming cycle greatly improves sample flushing allowing solvent-free cleaing
  • User interface options include an instrument calibration routine, configuration of test cycles, and multiple data output options including save, print and export for LIMS capture

Dimensions (W × D × H) Unit: 33.3 cm × 64.4 cm × 71.1 cm (13.1 in × 25.4 in × 28 in)
Waste receiver: 26.7 cm × 34.3 cm × 38.9 cm (10.5 in × 13.5 in × 15.3 in)
Recirculating cooler: 24.9 cm × 50.0 cm × 59.9 cm (9.8 in × 19.7 in × 23.6 in)
Weight Unit: 46 kg (102 lb)
Waste receiver: 8.2 kg (18 lb)
Recirculating cooler: 39.1 kg (86 lb)
Shipping dimensions (W × D × H) Box 1: 88.9 cm × 88.9 cm × 88.9 cm (35 in × 35 in × 35 in)
Box 2 (recirculating cooler): 81.3 cm × 61.0 cm × 106.7 cm (32 in × 24 in × 42 in)
Shipping weight Box 1: 140.6 kg (310 lb)
Box 2 (recirculating cooler): 59.0 kg (130 lb)
Maximum thoughput Up to 6 samples per hour
Automated sample capacity 30
Viscosity range 900 mPa·s (cP) to 25000 mPa·s (cP)
Temperature range and accuracy/stability −35 °C to −5 °C, ± 0.05 °C
Minimum sample volume 40 mL
Operating conditions 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II; Pollution Degree 2
Electrical specifications 115 Vac, 60 Hz; 230 Vac, 50/60 Hz; 1000 watts power consumption
Compliance CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc, 60 sec.); ROHS

 

CCS-2100 Automated Cold-Cranking Simulator consists of a patented thermoelectrically-cooled rotor/stator with vacuum system and injection pump, a 30 position sample table, a temperature verification kit, an integrated controller, a waste receiver, a recirculating cooler, proprietary software and a set of CANNON CL viscosity standards. Computer sold separately. Description  Part #
115 Vac, 60 Hz 9728-E46
230 Vac, 50 Hz 9728-E47
230 Vac, 60 Hz 9728-E49

 

Bottle, 2 oz 38-400 Flint Graduated AC Round (288/case)

$0.00


Kit User Spare Parts CCS-2100 (for one year)

$2,620.19


Low Temperature Viscosity Standards (General, CCS and CMRV)


Calibrated Temperature Probe for CCS-2100

$397.17