Type: Total sulfur analyzer for petroleum products by energy-dispersive X-ray fluorescence method
Conforming Standards: ISO 8754, ASTM D4294, etc.
Number of Tests: One(1)
Sample Volume: 5-20ml
Measuring Range: 0-6.00wt%
Repeatability: (Typical)
|
Range
|
Std Deviation
|
Measuring Time
|
|
0.008 wt% max
|
5 ppm
|
300sec x 3times
|
|
0.05 wt% max
|
10 ppm
|
100sec x 3times
|
|
0.05 - 0.50 wt%
|
20 ppm
|
100sec x 3times
|
|
0.51 - 1.50 wt%
|
30 ppm
|
100sec x 3times
|
|
1.51 - 2.00 wt%
|
50 ppm
|
100sec x 3times
|
|
2.01 - 4.00 wt%
|
100 ppm
|
100sec x 3times
|
|
4.01 - 6.00 wt%
|
200 ppm
|
100sec x 3times
|
C/H Ratio Compensation:
Automatic with 0.003 wt% max of compensation error when a 1 wt% sample is tested.
Measuring Time:
10 - 300 sec with a 10 sec increment. 1-99 times.
Calibration:
- Automatic two point calibration
- Manual two point calibration
- Automatic multiple-point calibration (up to 10 points)
- Manual multiple-point calibration (up to 10 points)
X-Ray Source:
X-Ray tube rated 7KV - 150mA
Sample Cup: 2-piece Teflon cup with disposable Mylar film
Printer: Built-in impact dot matrix printer (W57mm)
Display: LCD (20 characters x 4 lines)
X-Ray Leakage: 0.6μSv/Hr or less on instrument surface
Operating Temperature/Humidity:
10-30°C with a temperature drift within 10°C in a day.
RH to be within 80%.
Power Supply:
DC 12V (with the provided AC adapter or optional 12VDC cord with lighter plug)
Safety: Interlock mechanism against accidental X-ray leakage
Dimensions & Weight:
Test head: 420mm W x 350mm D x140mm H / 12.5kg
Power supply: 80mm W x 160mm D x 60mm H / 1.0kg
Standard Accessories:
| 1 |
AC Adapter |
1 pc |
| 2 |
Teflon sample cells |
1 box |
| 3 |
Sample cell assembling jig |
1 set |
| 4 |
Sample cell disassembling jig |
1 pc |
| 5 |
Sample cell stand |
1 pc |
| 6 |
Mylar film (200mm x 1,000m) |
1 roll |
| 7 |
Printer roll paper |
3 rolls |
| 8 |
Printer ink ribbon |
1 pc |
Optional Accessories:
| 1 |
12VDC cord with lighter plug (3.0m) |
| 2 |
RS-232C port (factory option) |
| 3
|
Certified Reference Material
Sulfur in gas oil: 100, 200, 500, and 800 ppm level
Sulfur in residual fuel oil: 0, 0.1, 0.2, 0.5, 1, 2, 3, and 4 wt% level
For a traceable calibration, the CRMs are necessary |
Suggested Spares For 2 Years:
| 1 |
Teflon sample cells (5 pcs/box) |
1 box |
| 2 |
Mylar film (200mm x 1,000m) |
1 roll |
| 3 |
Printer roll paper (5 rolls/pk) |
5 pks |
| 4 |
Printer ink ribbon |
5 pcs |