Conforming Standards:
ASTM D56 etc.
Measuring Range:
Ambient to 95oC (for sub-ambient, consult TANAKA)
Test Mode:
ASTM D56, Rapid, Special (Fast Search),
User’s Custom
Display:
Fluorescent display tube (VFD Module)
Temperature Sensor:
PT-100 in stainless steel sheath
Flash Detector:
CRC Thermocouple
Heater:
Plate type heater: 250W@100V or 250W@220V
Ignition Source:
Gas ignition with automatic lighting or electric ignition, interchangeable
Cooling After Test:
By replacing bath liquid
Barometric Correction:
By entering the barometric pressure reading through the control panel, or by optional barometric pressure sensor
Data Output:
RS-232C 1 channel (for PC or Optional Printer)
Safety Mechanism:
Automatically shuts off and the problem is reported by buzzer and display, in case:
| 1 |
EFP+10oC or at 95oC is reached |
| 2 |
Temperature sensor is found defective |
| 3 |
Flash detector is found defective |
| 4 |
Thermofuse is blown |
| 5 |
Electric Ignitor is blown |
| 6 |
Built-in battery is found drained |
| 7 |
Test cover is not set in place |
| 8 |
Flash detector detects continuous burning |
| 9 |
Control computer crashes (no display) |
Dimensions & Weight:
Control unit: 230W x 455D x 110H (mm), 6.0kg
Test unit: 230W x 480D x 275H (mm), 15.5kg
Installation Site:
Ambient temperature: 0 to 40oC
RH: Less than 90%(no condensation)
Utility:
AC100/120V or 220/240V 50/60Hz (set at the factory) 0.5kW(max.)
GAS SUPPLY (When Gas Ignition is used):
LP gas or natural gas (Max. pressure<9.8kPa)
Standard Accessories:
| 1 |
Test Cup |
1 pc |
| 2 |
O-Ring for Test Cup G-45 |
1 pc |
| 3 |
Power Connecting Cable, 0.6m |
1 pc |
| 4 |
Signal Connecting Cable, 0.6m |
1 pc |
| 5 |
AC Power Cord, 2.5m (<AC125V) or 3.0m (>AC200V) |
1 pc |
| 6 |
Spare Thermofuse |
3 pcs |
| 7 |
Insulation tube for Thermofuse, 0.1m |
1 pc |
| 8 |
Gas Hose 9x16x1500mm |
1 pc |
| 9 |
Gas Hose Band |
2 pcs |
| 10 |
Braided Hose for Water Supply 10x16x1500mm |
1 pc |
| 11 |
Wire Band |
4 pcs |
| 12 |
Strainer |
1 pc |
| 13 |
Drain Hose 15x19x1500mm |
1 pc |
| 14 |
Drain Hose Band |
1 pc |
| 15 |
Dripping Plate |
1 pc |
| 16 |
Wind Screen |
1 pc |
| 17 |
Instruction Manual |
1 pc |
| 18 |
oC/oF Display |
1 pc |
| 19 |
Data Storage Software (requires built in clock) |
1 pc |
Suggested Spares for 2 Years:
| 1 |
Test Cup |
1 pc |
| 2 |
O-Ring G-45 |
1 pc |
| 3 |
Electric Igniter |
2 pcs |
| 4 |
Thermofuse (pack of 5) |
1 pk |
| 5 |
Temperature Sensor |
1 pc |
| 6 |
Flash Detector |
1 pc |
Optional Accessories:
| 1 |
Power Connecting Cable, 3m |
| 2 |
Signal Connecting Cable, 3m |
| 3 |
Barometric Pressure Sensor |
| 4 |
Built-in Clock Board (for time/date verification) |
| 5 |
Printer, BS-80TSL (w/ AC Adapter & Connecting Cable) |
| 6 |
Changeover Unit, Model: CHG-7 |
| 7 |
Chiller Unit for sub-ambient Test (Use Altered Test Unit) |