HomeAbout CannonNewsClassroomFree CatalogPoliciesContact UsJobsView CartMini-Rotary ViscometerGlass ViscometersminiAV Automatic ViscometerStandardsCAV2000 Series Automatic Viscometers

ACO-7 Automated Cleveland Open Cup Flash Point Tester

Model ACO-7 is the latest version of the automated Cleveland Open Cup(COC) flash point tester from Tanaka. When this seventh generation model was designed after 35 years of providing automated flash point testers to the petroleum industry, ease of operation and maintenance was sought.

Model ACO-7 Asphalt Version is a new addition to the latest version of the automated Cleveland Open Cup(COC) flash point tester from Tanaka. While maintaining the features of the regular ACO-7 flash point tester, a unique skimmer was added to automatically remove the surface skin that forms on the sample when bitumen or some other samples are tested. Surface skin removal has been known to improve the precision and the safety of the COC flash point testing of such samples, and some test methods including JIS recommend it.



While data processing functions such as data storage and statistical analysis can be added to the automated flash point testers, these functions were excluded from the standard specification, since it was found that very few users would use these functions and the addition of such extra functions could make the operation of the regular flash point determination complicated.

Single Action Setting: The flash detector rings and the temperature sensor are permanently mounted on a swing-arm assembly, which allows easy setting of the specimen.

Easy Operation: Select a test mode and enter the expected flash point; while the instrument executes the test, you are free to do other lab work. The tester follows the exact procedures prescribed in the test method, and the completion of the test cycle is signaled by beeps. The test result is brightly shown on the VFD module.

Auto Fire Point Detection: In addition to the flash point, fire point can be determined automatically. The fire point is detected when the double ionization rings detect 5 sec of continuous burning.

Auto Fire Containment: ACO-7 is equipped with a fire containment lid which activates automatically to help extinguish the fire when ACO-7 detects a fire point or specimen is burning heavily.

Modular Design: The tester consists of a computer control unit and a test head unit, allowing remote placing of the test head unit up to 3 meters away. An optional Changeover unit can be purchased to attach the control unit to another type of flash point test head unit (APM-7 or ATG-7). The control unit is designed to control any of the three different flash point test head units, one at a time.

Specifications

Conforming Standards:
ISO 2592, ASTM D92, IP 36, etc.

Measuring Standards:
Ambient to 400 oC

Test Mode:
ASTM D92 Flash + Fire, Special (Fast Search),
User’s Custom

Display:
Fluorescent display tube (VFD Module)

Temperature Sensor:
PT-100 in stainless steel sheath

Flash Detector:
Ionization Ring

Heater:
Nichrome Coil heater: 800W@100V or 800W@220V

Ignition Source:
Gas ignition with automatic lighting

Skimming Mechanism (Asphalt Version Only):
Teflon paddle rotates to skim the surface of the specimen. Skimming start temp and interval are programmable. When not in use, the paddle can be removed.

Cooling Device:
Forced air cooling by sirocco fan

Barometric Correction:
By entering the barometric pressure reading through the control panel, or by an optional barometric pressure sensor

Data Output:
RS-232C 1 channel (for PC or Optional Printer)

Safety Mechanism:
Fire containment lid which activates automatically to help extinguish the fire when the ACO-7 detects a fire point

Automatically shuts off and the problem is reported by buzzer and display, in case:

    1 EFP+30oC or at 400oC is reached
    2 Temperature sensor is found defective
    3 Flash detector is found defective
    4 Thermofuse is blown (Also fire containment lid activates)
    5 Built-in battery is found drained
    6 Arm is not set in place
    7 Control computer crashes (no display)

Dimensions & Weight:
Control unit: 230W x 455D x 110H (mm), 6.0kg
Test unit: 230W x 450D x 275H (mm), 14.0kg
Test unit: 230W x 450D x 345H (mm), 14.5kg (Asphalt Version)

Installation Site:
Ambient temperature: 0 to 40oC
RH: Less than 90%(no condensation)

Utility:
AC100/120V or 220/240V 50/60Hz (set at the factory) 1.0kW(max.)
GAS SUPPLY
LP gas or natural gas (Max. pressure<9.8kPa)

Standard Accessories:

    1 Test Cup Assembly 1 pc
    2 Power Connecting Cable, 0.6m 1 pc
    3 Signal Connecting Cable, 0.6m 1 pc
    4 AC Power Cord, 2.5m (<AC125V) or 3.0m (>AC200V) 1 pc
    5 Thermal Insulation (Heating Plate) 1 pc
    6 Spare Thermofuse 3 pcs
    7 Insulation tube for Thermofuse, 0.1m 1 pc
    8 Spare Teflon Paddle (Asphalt Version Only) 1 pc
    9 Gas Hose 9x16x1500mm 1 pc
    10 Gas Hose Band 2 pcs
    11 Instruction Manual 1 pc
    12 oC/oF Display 1 pc
    13 Data Storage Software (requires built in clock) 1 pc

Suggested Spares for 2 Years:

    1 Test Cup Assembly 1 pc
    2 Igniter Top (Ignition Filament) 2 pcs
    3 Thermofuse (pack of 5) 1 pk
    4 Temperature Sensor 1 pc
    5 Inner Detector Ring 1 pc
    6 Outer Detector Ring 1 pc
    7 Insulating Collar for Flash Detector 2 pcs
    8 Spare Teflon Paddles (Asphalt Version Only) 5 pcs
    9 Nichrom Heater Coil 800W@100V or 800W@220V 1 pc

Optional Accessories:

    1 Mercury Thermometer Holder for Calibration
    2 Power Connecting Cable, 3m
    3 Signal Connecting Cable, 3m
    4 Barometric Pressure Sensor
    5 Built-in Clock Board (for time/date verification)
    6 Printer, BS-80TSL (w/ AC Adapter & Connecting Cable)
    7 Changeover Unit, Model: CHG-7